Item# DC6100

Download Product Brochure
DC6100 Remote Scanner Mainframe. Select a Mainframe Capacity (30 or 60 Channels), an AC or DC Power Supply, and Select either Control Unit or Channel Expansion Unit configuration. NOTE: Each Control Unit connects up to four (4) Channel Expansion Units using Synchronization Cables. Four (4) Control Units connect to a PC Host to activate 1200 channels of data collection. Accessories Include: Screw Drive for Connection, Terminal Connector and PC Control Software.
For configuration assistance and pricing, please contact SOLTEC
|
|
FeaturesSpecificationsMore Information
Great Extendibility of Channels
From small-scale measurement in the Lab (10 channels) to large-scale measurement in the Plant (up to 1,200 channels).
Multi-Scan Function
A single unit is capable of setting three (3) scanning rates. It allows effective use of internal memory or compact flash card by selecting optimal scanning time for each object to be measured.
High-Speed Sampling
Max. 10ms sampling when using 30 Channels.
Max. 100ms sampling when using 1200 Channels.
Input Scanner
Five (5) types of input units are available: 10-Channel Temperature/Voltage, Universal, Strain, 30-Ch Temperature/Voltage and DI/ALM units.
Various Measurement Methods
Effective for repetitive testing. Selectable from Constant Interval, Variable Interval and Step-Trigger measurement modes.
Online Access Function
Internet accessibility with options to control data settings, data monitoring, file management, time setting and more.
|
| DC6100 Mainframe |
| Item |
Model |
Note |
| Remote Scanner |
DC6100 |
□ |
□ |
□ |
Accesories:
Screw Drive for Conection, Terminal Connector and PC Control Software
|
| Selection of Unit (to be installed at the factory) |
Control Unit (DC61-101) |
C |
|
Select C or E |
| Channel Expansion Unit (DC61-103) |
E |
|
| AC Power Unit (DC61-102) |
A |
|
Select A or D |
| DC Power Unit (DC61-104) |
D |
|
| 30 Channel Mainframe (DC61-151) |
3 |
Select 3 or 6 |
| 60 Channel Mainframe (DC61-152) |
6 |
Input Unit |
| Item |
Model |
Note |
| 10-Ch Temp/Voltage Unit |
DC61-201 |
10 Channels, Screw Connector |
| Universal Unit |
DC61-202 |
10 Channeld, Screw Connector |
| Strain Unit |
DC61-203 |
10 Channels, Bayonet Miniture Circular Connector Block |
| DI/ALM Unit |
DC61-204 |
Digital/Pulse Input 10 Channels, Alarm output 4 Channels, Screw Connector |
| 30 Ch Temp Voltage Unit |
DC61-205 |
30 Channe;s, Screw Connector |
| Options |
| Item |
Model |
Notes |
| Blank PaneCarrying Kit |
DC61-210 |
Blank Panel to cover open input unit slot |
| Carrying Kit |
DC61-408 |
Handle, Protector |
| DIN Rail Mount Kit |
DC61-409 |
DIN Rail Mount Attachment |
| Rack Mount Kit |
DC61-410 |
Rack Mount Bracket for JIS Standard |
| 19-Inch Rack Mount Kit |
DC61-414 |
19-Inch Rack Mount Bracket for EIA Standard |
| Dust Cover for 30-Ch |
DC61-411 |
Dust Cover with Protector |
| Dust Cover for 60-Ch |
DC61-412 |
Dust Cover with Protector |
| Temp/Voltage Terminal Block Unit |
DC61-331 |
Input Terminal Block |
| Universal Terminal Block Unit |
DC61-333 |
Input Terminal Block |
DI/ALM Terminal Block Unit |
DC61-336 |
Input Terminal Block |
Synchronized Cable for Channel Expansion |
DC61-401 |
2m |
| Synchronized Cable for Channel Expansion |
DC61-402 |
5m |
| Synchronized Cable for Channel Expansion |
DC61-403 |
10m |
| Conversion Cable for Strain Input |
DC61-411 |
0.6m |
| Item |
Model |
Specifications |
| Control Unit |
DC61-101 |
Required unit for a master device
PC Connection via either Ethernet (100BASE-TX) or USB (Version 2.0)
Built-in syncchronized measurement function, internal memory: 1 GB
Compact flash card slot
Function to turn on/off expansion unit
Can connect up to 4 slave units by exclusive-use connectioncable to expand measuring channels
|
| Channel Expansion Unit |
DC61-103 |
Required unit for slave units
For connecting high low order devices uding exclusive-use cable to expand measuring channels, D-Sub 9 pin connnector x2
|
| AC Power Unit |
DC61-102 |
AC Power 90 to 264VAC (50/60Hz) |
| DC Power Unit |
DC61-104 |
DC Power10.5 to 264 VDC |
| 30-Ch Mainframe |
DC61-151 |
Slots to power unit, control unit, or channel expansion unit and input units (3 units) |
| 60-Ch Mainframe |
DC61-152 |
Slots to power unit, control unit, or channel expansion unit and input units (6 units) |
| 10-Ch Temperature/Voltage Unit |
DC61-201 |
Channel
Input Type
|
10 Channels
DC Voltage: ±20mV, 200mV, 2V, 20V, 50V
Thermocouple: R, S, B, K, E, J, T, N, W, PR, KpAu7Fe
RTD: Pt100/JPt100, 3-line type
Strain: ±20,000µ∑, ±200,000µ∑
Contact: ON at 2kΩ or below, OFF at 5kΩ Or over |
| Universal Unit |
DC61-202 |
Channel
Input Type
|
10 Channels
DC Voltage: ±20mV, 200mV, 2V, 20V, 50V
Thermocouple: R, S, B, K, E, J, T, N, W, PR, KpAu7Fe
Strain: ±20,000µ∑, ±200,000µ∑
Contact: ON at 2kΩ or below, OFF at 5kΩ Or over |
| Strain Unit |
DC61-203 |
Channel
Input Type
|
10 Channels
DC Voltage: ±20mV, 200mV, 2V, 20V, 50V
Strain: ±20,000µ∑, ±200,000µ∑
Contact: ON at 2kΩ or below, OFF at 5kΩ Or over
|
| DVALM Module |
DC61-204 |
Channel
Input Type
Output Type
|
Digital/Puse Input:10 Channels Alarm output: 4 Channels
Pulse input and digital input (H/L) selectable by channel
4-point relay output (Select Contract A or B by connection terminal)
|
|
|