FeaturesSpecificationsMore Information
- Large Color Screen Display. A large 12.1" full color LCD touch screen displays set-up information and acquired data in real time or from memory. View before print or print while viewing data.
- Unique Pen Recorder Mode. Clear graphic, touch to set, amplifier control images eliminate complicated recorder setup. Displayed data signals appear to originate from graphic writing pen nib image while data is recording.
- Enhanced Ease of Stored Data Access. Finding and displaying recorded data has never been easier. Many search features for finding specific data points within a large data file are included. A Thumbnail Bar displays all recorded channel data, while a Jump search function finds max/min points, time, etc.
- Direct Sensor Input. Sensors may be directly connected to the wide selection of plug-in signal conditioning amplifiers to measure AC/DC Voltage, AC/DC Strain, Temperature, Vibration, Pressure, Rotation Pulses, Frequency, etc.
- Long-Term Recording. A built-in 60GB HDD provides long-term, high speed data recording for over 180 days on 16-Channels with a 10mS simultaneous sampling rate! (90 days on 32-Channels)!
- Standard LAN & USB Ports. LAN (100BASE-T) for data communication and USB for external storage devices (MO/USB memories) are standard interfaces.
- Built-In Chart Printer. A high resolution, thermal array printer with wide chart paper allows high-speed real time chart printing at 100mm/sec or prints stored files, and portions thereof, on demand.

|
Basic Specifications:
Display
|
12.1" TFT Color LCD (1024 x 768 pixels)
|
Channel
|
TA220-2300
|
16 Channels (8 Slots) + Optional 16 Channel Digital Inputs
|
TA220-2800
|
32 Channels (16 Slots) + Built-in 16 Channel Digital Inputs
|
Sampling Speed
|
TA220-2300
|
MAX 1µs (1MS/s)
|
TA220-2800
|
MAX 2µs (500kS/s)
|
Printer
|
Printing Method
|
Thermal
|
Paper
|
Width |
219.5mm
|
| Length |
Roll: 30m
|
| Z-Fold: 210m (requires optional adapter) |
Effective Recording Width
|
1 Division (200mm FS) to 16 Divisions (10mm FS). Number of Divisons and Printing Width can be changed.
|
Channel Discrimination
|
Printout includes Channel Number, Waveform and Scale
|
Grid Pattern
|
Standard (10mm, 5mm), 10mm, 5mm, No Grid
|
Battery Backup
|
Clock, Setting Value: Approx. 3 to 5 years
|
Storage Device
|
60GB Hard Disk Drive (HDD), Built-In Standard
|
MO or USB Drive via USB Port
|
Interface
|
Standard
|
LAN (10/100 BASE-T), USB
|
Optional
|
RS-232C, Remote Terminal
|
Operating Environment
|
Temperature
|
5°C to 40°C
|
Humidity
|
35 to 80%RH (non-condensing)
|
Power Supply
|
90 to 264VAC, Frequency 50 to 60Hz
|
Power Consumption
|
100VA (typical): with TA220-AP11-101 x 8 units*
|
Dimensions
|
TA220-2300
|
Approx. 369.5(W) x 150.5(H) x 301(D) mm (excluding projections)
|
TA220-2800
|
Approx. 400(W) x 270(H) x 380(D) mm (excluding projections
|
Weight
|
8.0kg or less (TA220-2300 main body only)
|
8.7kg or less (main body with TA220-AP11-101 x 4 units)
|
*Paper Feed Speed: 20mm/s, Input Signal: 10Hz
Communication & Storage Specifications:
HDD
|
Function
|
Set Conditions of Main Unit and Record/Display Measured Data
|
Capacity
|
60GB
|
LAN
|
Function
|
Control with Communication Command, Windows and File Sharing with Windows PC
|
Standard
|
10/100BASE-T
|
USB
|
Function
|
Data Saving on Storage Device by USB Connection
|
Standard
|
1.1
|
Storage Devices
|
MO/USB Drive and HDD Devices specified by SOLTEC
|
Trigger Specifications:
Trigger Mode
|
OR, AND, WINDOW, OFF |
Trigger Source
|
Input Signal, Manual Trigger, External Trigger |
Trigger Settings
|
Amps other than Event Amp
|
Trigger Slope |
OR, AND or ↑ or ↓ WINDOW OUT or IN
|
| Level Setting |
To be Set with Physical Values (e.g. Voltage)
|
|
Event Unit (Optional)
Event Amp (TA220-AP11-105)
|
State Settings |
H, L, or OFF can be set for each Input (without Trigger Slope)
|
| Trigger Setting |
AND or OR of State Setting Conditions of Inputs from 1 to 8
|
Trigger Related Functions
|
Trigger Output
|
Print/Display Signal when Trigger Conditions are met |
| (0 to 5V Signal Active LOW; Pulse Width, Approx. 10ms) |
Pre-Trigger
|
Pre-Trigger: 0 to 100% (1% Step) |
Trigger Mark
|
Record Trigger Point with an Arrow (↓). |
| Also prints Year, Date and Time the Trigger Occurred. |
Trigger Filter
|
1 to 65534 Samples |
Measuring Mode (Acquisition/Recording) Specifications:
Waveform Printing
|
Function
|
Paper Printout of Input Signal
|
Measurement Start Commands
|
Press Start Key, Trigger Detection or Preset Time
|
| Interval Recording Available |
Paper Speed
|
100mm/s to 1mm/min (User Setting, External Synchronization Enabled)
|
Frequency Response
|
DC to 100kHz (Sampling: 10 points/cycle). Varies by Input Unit.
|
Printing Density
|
Voltage Axis
|
8 dots/mm
|
Time Axis
|
80 dots/min (at 25mm/s)
|
Printing Length
|
Continuous
|
Data Recording
|
Function
|
Real-Time Recording of Measured Data on HDD
|
Recordable Size
|
55GB
|
Sampling Speed
|
1µs (1CH), 5µs (8CH), 10µs (16CH) max
|
Recording Method
|
Normal or Repeated Recording during Present Time
|
|
Waveform Printing
(See Pen Recorder)
|
Function
|
Printout Input Signal Data on Recording Paper (Waveform)
|
Measurement Starting Operation
|
ON/OFF of Printout to Recording Paper while HD Recording
|
Paper Speed
|
100mm/s to 1mm/min (User Setting, External Synchronization Enabled)
|
Frequency Response
|
DC to 100kHz (Sampling: 10 points/cycle). Varies by Input Units
|
Printing Density
|
Voltage Axis
|
8 dots/mm
|
Time Axis
|
80 dots/mm (at 25mm/s)
|
Printing Length
|
Continuous
|
Data Recording
|
Function
|
Records Measured Data to Internal Memory
|
Measurement Operation
|
Once, Repeat, and Endless
|
Memory Capacity
|
32MW. When using 16-Channels, 2MW/Ch
|
Memory Division
|
1, 2, 4, 8, 16, 32, 64 and 128 Divisions
|
Sampling Speed
|
1µs (1MS/s) to 100s (1S/s), User Setting, External Synchronization Enabled
|
Waveform Recording
|
Function
|
Print Memory Data onto Recording Paper (Waveform)
|
Printing Density
|
Voltage Axis
|
8 dots/mm
|
Time Axis
|
10 dots/mm
|
Copy Magnification
|
x5 to x1/1000
|
Memory Filing
|
Data is saved to the Memory Device in Binary or CSV format
|
Data Backup
|
Memory Back-up with HDD
|
(Data saved in a Specified Area of HDD at Shutdown)
|
Save/Copy Area
|
Copy between Cursors or Copy Area around Trigger Point 1% FS steps up to 100% FS
|
Function
|
Steady-State and Transient Events can be Recorded Simultaneously on HDD, Memory and/or Recording Paper
|
Pen Recorder
|
Refer to Pen Recorder Specifications
|
HDD Recorder
|
Refer to HDD Recorder Specifications
|
Memory Recorder
|
Refer to Memory Recorder Specifications (Waveform Printing Not Available)
|
Function
|
ON/OFF of Locus Enable (Pen Up & Down)
|
Input Signal Monitor, Freeze, Copy and X-Y Display During Data Recording Available
|
Axis Setting
|
X Axis
|
1 Channel
|
Y Axis
|
3 Channels
|
Measuring Speed
|
1ms to 1s
|
Data Recording
|
Function
|
Record all Input Signals to HDD
|
Recordable Size
|
55GB Max
|
Recording Method
|
Standard
|
Waveform Printing
|
Function
|
Printout Display Waveforms
|
X Axis: One (1) 1 Channel
|
Y Axis: Three (3) 3 Channels
|
A4 Format
|
Resolution
|
Printout
|
1600 x 1600 dots
|
TFT Display
|
650 x 650 dots |
Measured Data Display (Replay Monitor Specifications):
Function
|
Display Recorded Data when X-T or X-Y "Replay" Mode is Selected
|
Available Measuring Modes
|
Playback Stored Data Independent of Recording Mode
|
Y-T Display
|
Waveform Division
|
1 to 16 Divisions
|
Display Magnification
|
x100 to x1/10,000
|
Thumbnail Function
|
Display Data of the Selected Channel on Thumbnail Bar
|
Numeric Display
|
Numeric Value, Cursor Value, Numeric + Cursor Values
|
Search Functions
|
Search by Cursor, Time, Address, Max/Min and Event
|
X-Y Display
|
Channels Allowed
|
X Axis
|
Displays up to One (1) Channel
|
Y Axis
|
Displays up to Three (3) Channels
|
Data Input
|
To File and/or Print on Chart Paper
|
Output File Format
|
Binary or CSV
|
Other Specifcations:
Printer
|
Data Information
|
Measuring Mode, Year/Month/Day, Measrement Start Time, Data Number, Trigger Condtions (Trigger Point, Trigger Date, Trigger Time), Sampling Speed, Paper Speed, Time Axis can be printed with Waveforms. ON/OFF Selectable.
|
Channel Information
|
Print Input Unit Settings when Saved. ON/OFF Selectable.
|
Mark Print
|
Prints Event Mark and Time on Chart
|
List Print
|
Printout of Condition Settings (Sampling Speed, Detailed Amp Settings, etc.) Available
|
Screen Copy
|
Print Screen Image
|
Line Width for Printing
|
Select Base Line Thickness for each Channel (1, 2, 3 or 4 dots)
|
Auto Function
|
Function
|
Sampling Speed and Input Range can be Set Automatically by Using "Auto" Mode
|
Sampling Speed
|
Auto Adjusts Sampling Speed of Active Shannels based on Input Signal
|
Input Range
|
Auto Range Adjust of all Active Channels
|
Timer Function
|
Start/End Time and Interval can be set
|
CSV Conversion
|
Batch Conversion of Multiple Memory Blocks or Files
|
Screen Image Saving
|
Save Screen Image on HDD in BMP Format
|
Save/Readout of Settings
|
Save up to four (4) Settings of Input and Main Unit Setting Conditions on the HDD
|
Keylock Function
|
Prevents Operational Error (Password Protected)
|
Physical Value Conversion
|
Physical Conversion of Input Signals, Full Scale Change on Display, Registration of Units
|
Event Unit Waveform Display
|
Display position of EVENT AMP and EVENT UNIT Information (optional) on Graphic Output can be Changed. Set Display Position and Spacing.
|
Optional Units:
- Remote Unit (TA220-RA23-112)
Function
|
Start, Stop, Mark Print, Paper Feed is possible via External Signal. Synchronization Pulse Input. Output Error Signal, Input UPS Protect Signal.
|
Cable
|
1.5m, I/O Connector (28-pin) and Bare Wire
|
Weight
|
60g or less
|
- Event Unit (TA220-RA23-113)
Function
|
Inputs Logic Signals Directly into the Main Unit (Independent from Plug-In Amps)
|
Number of Signals
|
16
|
Signal Level
|
0 to 5V, Voltage Only Input (no contact)
|
Cables
|
1.5m, I/O Connector (34-pin) and Bare Wire
|
Weight
|
60g or less
|
- RS-232C Unit (TA220-RA23-114)
Standard
|
JIS X5101 (former C6351) Compliant
|
Transfer Speed
|
38400, 19200, 9600, 4800 or 2400bps
|
Connector
|
D-Sub (9-pin) Connector
|
Function
|
Shutdown
|
UPS Enabled Shutdown
|
Remote Control
|
PC control via RS-232
|
Weight
|
50g or less
|
- AC Bridge Power Supply Unit (TA220-RA23-116)
Function
|
Bridge AC Power Source for 2-Ch Carrier Strain Amp TA220-AP11-104. Powers up to eight (8) each TA220-AP11-104 Amp Units.
|
| Power Voltage |
2Vrms |
Carrier Wave
|
Sine Wave 5kHz
|
Synchronization
|
Synchronizes with Bridge Power Units Installed in other TA220-2300MF Units.
|
Weight
|
60g or less
|
|
Related items you may need...
|